I-E67-TB82FF-2A February 13, 2002
20
UTILITY MODE
The Utility Mode of Operation provides access to
powerful functions not used during normal operating
conditions. These functions have been separated into
two categories: Factory and User. Factory functions
are strictly reserved for factory personnel. User
functions include Reset Configuration to default
settings, Reset Security password, Reset All
parameters to default settings, Software Reboot, FF
Identification and Damping functions.
The User States of Operation can be accessed using
the hidden fifth key located top, center of the keypad
above the display window (see Figure 2). Once the
hidden key has been pressed, the secondary display
will have the prompt ‘USER’ shown. The SELECT
key provides access to the User States.
USER
TB.OOS?
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
YES
RST.CON
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
RST.ALL
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
RST.SFT
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
12.3
12.3
12.3
12.3
RST.SFT
RST.ALL
RST.CON
MEASURE
RST.CON
RETURN
RSTALL
RETURN
RST.SFT
RETURN
SELECT
SELECT
SELECT
RST.SEC
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
12.3
RST.SEC
RST.SEC
RETURN
SELECT
NEXT
NEXT
FF.ID
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
12.3
SELECT
FF.ID
FF.ID
RETURN
NO
DAMPNG
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
12.3
SELECT
DAMPNG
DAMPNG
RETURN
mS/cm
mS/cm
mS/cm
mS/cm
mS/cm
mS/cm
mS/cm
Figure 29 – Screen Flow Diagram for User States
of Operation
Reset Configuration State
The Reset Configuration State returns the
configuration to factory default settings. If the
Configure Mode has been password secured, the
same password will be required to perform a reset to
the transmitter’s configuration. See the Configuration
Worksheet at the end of this manual for software
default settings.
Reset Security State
The Reset Security State returns the security to
factory default settings. The factory default is security
OFF for all applicable modes (i.e., Security, Calibrate,
and Configure). To reset the security, the password
732 must be entered when requested by the
transmitter.
Reset All State
The Reset All State returns all transmitter parameters
back to factory defaults. This includes calibration,
configuration, security, and secondary display values.
To reset all transmitter parameters, the password 255
must be entered when requested by the transmitter.
Reset Software State
The Soft Boot State initiates a software reset. A
software reset repeats the boot-up and self-test
process. All programmable instrument parameters
are unaffected by this function.
Caution
Since the Utility Mode contains functions that can have
a dramatic effect on the proper function of the
transmitter, the Transducer Block should be put Out Of
Service until all transmitter parameters have been
properly set. For convenience, the Transducer Block
can be put Out Of Service at the local HMI; however,
once the transmitter is returned to the Measure Mode of
Operation using the Exit to MEASURE key, the
transmitter will automatically return the Transducer
Block into its previous state.
Note:
All user specific information will be lost once a Reset All
or Reset Configuration has been initiated. Before
initiating these reset functions, record configuration data
to make reconfiguration quicker
.