I-E67-TB82FF-2A February 13, 2002
14
Any changes to the transmitter configuration must be
saved. Pressing the Exit to MEASURE key prompts
the user to “SAVE?” their changes. Pressing the YES
key saves the new configuration and returns the
transmitter to the Measure Mode. The NO key
discards the changes and returns the transmitter to
the Measure Mode. Upon exiting from the Modify
state, the TB mode is returned from Out-Of-Service to
its previously set state.
The Modify Configure State contains all the available
settings that establish the functionality of the
transmitter. Upon receipt of the transmitter, the
default configuration (unless otherwise specified by
the customer when ordering the transmitter) will be
active once the transmitter has been powered. See
the Configuration Data Sheet at the end of this
manual for default configuration settings. Before
installing the transmitter, the configuration should be
modified to reflect the final installed application. The
Table below describes each of the Modify Configure
States and their function.
State Function
ANALYZR Used to define the type of analyzer. Choices
include Conductivity and Concentration.
TMP.SNS Used to define the type of temperature
sensor. Choices include None, Pt100, 3k
Balco, and 4.75 kohm Standard Network.
TC.TYPE Used to define the type of temperature
compensation. Choices include Manual
(0.1N KCl based), and Automatic, which can
be set to one of the following: Standard KCl
(0.1N KCl based), Temperature Coefficient
(%/
°
C), 0 to 15% NaOH, 0 to 20% NaCl, 0 to
18% HCl, 0 to 20% H
2
SO
4
, .Pure Water
Neutral, Pure Water Acid, Pure Water Base,
and User-Defined.
DIAGS Used to set the sensor diagnostics ON or
OFF.
TC.TYPE
RETURN
CONFIG
EXIT
ANALZ R
RETURN
ANALZR
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SELECT
NEXT
ANALZ R
TMP.SNS
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SELECT
NEXT
TC.TYPE
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SELECT
NEXT
TMPSNS
TC.TYPE
TMPSNS
RETURN
mS/cm
12.3
mS/cm
12.3
mS/cm
12.3
DIAGS
RETURN
DIAGS
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SELECT
NEXT
DIAGS
mS/cm
12.3
TB.OOS?
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NO
YES
mS/cm
12.3
Figure 17 – Screen Flow Diagram for Modify
Configure States of Operation
Analyzer State
The Analyzer State sets the type of measurement
(i.e., Process Value) and sensor type. The
measurements are conductivity and concentration.
The Conductivity setting displays the Process Value
using the engineering units of either mS/cm or µS/cm.
The Concentration setting converts the measured
temperature compensated conductivity to a solute
weight percentage using either one of the canned
concentration algorithms or a user-defined curve fit.
Concentration units are set to percent or can be can
separately defined for a user-defined curve fit.
ANALZR
mS/cm
12.3
COND
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
mS/cm
12.3
CONCEN
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
CONFIG
EXIT
Last Selected
USDEFN
ENTER
ENTER
ANALZR
RETURN
mS/cm
12.3
A GRP
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
ENTER
mS/cm
12.3
B GRP
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
ENTER
mS/cm
12.3
C GRP
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
ENTER
CONCEN
GRPA
FOR CONCEN
SELECTION ONLY.
FOR CONCEN
SELECTION ONLY.
FOR
CONCEN
SELECTION
ONLY.
FOR
CONCEN
SELECTION
ONLY.
Figure 18 – Screen Flow Diagram for Analyzer
States of Operation
The Conductivity Analyzer State contains three
choices for four-electrode sensor groups: A, B, and C.
The Sensor Group must be entered and correspond
with the conductivity sensor being used with the
transmitter. The measurement range for each sensor
group is listed in the Specification Section. Since
ABB four-electrode Conductivity Sensors have unique
cell constants that are defined by the physical
orientation, surface area of the electrodes, and the
measured solution voltage set by the electronics of
the transmitter, only ABB TB4 sensors are
recommended for use with the TB82EC FF
Transmitter.
The Concentration State converts conductivity values
to concentration units. This state applies
temperature-compensated conductivity
measurements to a pre-defined or user-defined